User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Fluke 5200A:AC Calibrator DATE: 28-Oct-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 4 NUMBER OF LINES: 36 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 HEAD INITIAL CONDITIONS 1.004 DISP Connect the UUT GUARD to its OUTPUT LO. 1.005 DISP Connect the UUT as follows: 1.005 DISP OUTPUT HI .......................... Datron 1281 HI 1.005 DISP OUTPUT LO .......................... Datron 1281 Lo 1.005 DISP SENSE HI ........................... Datron 1281 HI 1.005 DISP SENSE LO ........................... Datron 1281 LO 1.006 JMP 2.001 1.007 EVAL 2.001 CALL Fluke 5200A Sub 1 2.002 JMP 3.001 2.003 EVAL 3.001 CALL Fluke 5200A Sub 2 3.002 JMP 4.001 3.003 EVAL 4.001 CALL Fluke 5200A Sub 3 4.002 DISP Turn off the UUT. 4.003 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 15 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.